CVE-2017-14018
MEDIUM
4,8
Source: [email protected]
Attack Vector: physical
Attack Complexity: high
Privileges Required: none
User Interaction: none
Scope: unchanged
Confidentiality: none
Integrity: high
Availability: low
LOW
3,3
Source: [email protected]
Access Vector: local
Access Complexity: medium
Authentication: none
Confidentiality: none
Integrity: partial
Availability: partial
Description
AI Translation Available
An improper authentication issue was discovered in Johnson & Johnson Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017. The security authentication mechanism used between the Ethicon Endo-Surgery Generator Gen11 and single-patient use products can be bypassed, allowing for unauthorized devices to be connected to the generator, which could result in a loss of integrity or availability.
EPSS (Exploit Prediction Scoring System)
Trend Analysis
EPSS (Exploit Prediction Scoring System)
Prevede la probabilità di sfruttamento basata su intelligence sulle minacce e sulle caratteristiche della vulnerabilità.
EPSS Score
0,0014
Percentile
0,3th
Updated
EPSS Score Trend (Last 91 Days)
287
Improper Authentication
DraftCommon Consequences
Security Scopes Affected:
Integrity
Confidentiality
Availability
Access Control
Potential Impacts:
Read Application Data
Gain Privileges Or Assume Identity
Execute Unauthorized Code Or Commands
Applicable Platforms
Technologies:
ICS/OT, Not Technology-Specific, Web Based
Operating System
Endo-Surgery Generator Gen11 Firmware by Ethicon
CPE Identifier
View Detailed Analysis
cpe:2.3:o:ethicon:endo-surgery_generator_gen11_firmware:-:*:*:*:*:*:*:*
Common Platform Enumeration - Standardized vulnerability identification
https://ics-cert.us-cert.gov/advisories/ICSMA-17-332-01
http://www.securityfocus.com/bid/101978
https://ics-cert.us-cert.gov/advisories/ICSMA-17-332-01
http://www.securityfocus.com/bid/101978