Description

AI Translation Available

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

924

Improper Enforcement of Message Integrity During Transmission in a Communication Channel

Incomplete
Common Consequences
Security Scopes Affected:
Integrity Confidentiality
Potential Impacts:
Gain Privileges Or Assume Identity
Applicable Platforms
All platforms may be affected