CVE-2019-25719
HIGH
8,8
Source: [email protected]
Attack Vector: network
Attack Complexity: low
Privileges Required: none
User Interaction: none
Confidentiality: N/A
Integrity: N/A
Availability: N/A
HIGH
8,6
Source: [email protected]
Attack Vector: network
Attack Complexity: low
Privileges Required: none
User Interaction: none
Scope: unchanged
Confidentiality: low
Integrity: low
Availability: high
Description
AI Translation Available
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
924
Improper Enforcement of Message Integrity During Transmission in a Communication Channel
IncompleteCommon Consequences
Security Scopes Affected:
Integrity
Confidentiality
Potential Impacts:
Gain Privileges Or Assume Identity
Applicable Platforms
All platforms may be affected
https://static.draeger.com/security
https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-…