CVE-2019-25723
MEDIUM
6,3
Source: [email protected]
Attack Vector: network
Attack Complexity: high
Privileges Required: none
User Interaction: none
Confidentiality: N/A
Integrity: N/A
Availability: N/A
MEDIUM
4,0
Source: [email protected]
Attack Vector: network
Attack Complexity: high
Privileges Required: none
User Interaction: none
Scope: changed
Confidentiality: none
Integrity: none
Availability: low
Description
AI Translation Available
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
1286
Improper Validation of Syntactic Correctness of Input
IncompleteCommon Consequences
Security Scopes Affected:
Other
Potential Impacts:
Varies By Context
Applicable Platforms
All platforms may be affected
https://static.draeger.com/security
https://www.vulncheck.com/advisories/dr-ger-perseus-a500-dos-via-medibus-interf…