CVE-2019-25724

Published: Giu 02, 2026 Last Modified: Giu 02, 2026
ExploitDB:
Other exploit source:
Google Dorks:
HIGH 7,1
Attack Vector: adjacent
Attack Complexity: low
Privileges Required: none
User Interaction: none
Confidentiality: N/A
Integrity: N/A
Availability: N/A
MEDIUM 6,5
Attack Vector: adjacent_network
Attack Complexity: low
Privileges Required: none
User Interaction: none
Scope: unchanged
Confidentiality: none
Integrity: none
Availability: high

Description

AI Translation Available

Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

400

Uncontrolled Resource Consumption

Draft
Common Consequences
Security Scopes Affected:
Availability Access Control Other
Potential Impacts:
Dos: Crash, Exit, Or Restart Dos: Resource Consumption (Cpu) Dos: Resource Consumption (Memory) Dos: Resource Consumption (Other) Bypass Protection Mechanism Other
Applicable Platforms
Technologies: Not Technology-Specific, AI/ML
View CWE Details
https://static.draeger.com/security
https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-d…