CVE-2020-25187
HIGH
8,8
Source: [email protected]
Attack Vector: adjacent_network
Attack Complexity: low
Privileges Required: none
User Interaction: required
Scope: changed
Confidentiality: high
Integrity: high
Availability: high
HIGH
10,0
Source: [email protected]
Access Vector: network
Access Complexity: low
Authentication: none
Confidentiality: complete
Integrity: complete
Availability: complete
Description
AI Translation Available
Medtronic MyCareLink Smart 25000 is
vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device
EPSS (Exploit Prediction Scoring System)
Trend Analysis
EPSS (Exploit Prediction Scoring System)
Prevede la probabilità di sfruttamento basata su intelligence sulle minacce e sulle caratteristiche della vulnerabilità.
EPSS Score
0,0127
Percentile
0,8th
Updated
EPSS Score Trend (Last 91 Days)
122
Heap-based Buffer Overflow
DraftCommon Consequences
Security Scopes Affected:
Availability
Integrity
Confidentiality
Access Control
Other
Potential Impacts:
Dos: Crash, Exit, Or Restart
Dos: Resource Consumption (Cpu)
Dos: Resource Consumption (Memory)
Execute Unauthorized Code Or Commands
Bypass Protection Mechanism
Modify Memory
Other
Applicable Platforms
Languages:
C, C++, Memory-Unsafe
787
Out-of-bounds Write
DraftCommon Consequences
Security Scopes Affected:
Integrity
Availability
Other
Potential Impacts:
Modify Memory
Execute Unauthorized Code Or Commands
Dos: Crash, Exit, Or Restart
Unexpected State
Applicable Platforms
Languages:
Assembly, C, C++, Memory-Unsafe
Technologies:
ICS/OT
Operating System
Mycarelink Smart Model 25000 Firmware by Medtronic
CPE Identifier
View Detailed Analysis
cpe:2.3:o:medtronic:mycarelink_smart_model_25000_firmware:*:*:*:*:*:*:*:*
Common Platform Enumeration - Standardized vulnerability identification
https://us-cert.cisa.gov/ics/advisories/icsma-20-345-01
https://global.medtronic.com/xg-en/product-security/security-bulletins/mycareli…
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-345-01