CVE-2022-4992

Published: Giu 03, 2026 Last Modified: Giu 03, 2026
ExploitDB:
Other exploit source:
Google Dorks:
HIGH 8,8
Attack Vector: network
Attack Complexity: low
Privileges Required: none
User Interaction: none
Confidentiality: N/A
Integrity: N/A
Availability: N/A
HIGH 8,6
Attack Vector: network
Attack Complexity: low
Privileges Required: none
User Interaction: none
Scope: unchanged
Confidentiality: low
Integrity: low
Availability: high

Description

AI Translation Available

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

345

Insufficient Verification of Data Authenticity

Draft
Common Consequences
Security Scopes Affected:
Integrity Other
Potential Impacts:
Varies By Context Unexpected State
Applicable Platforms
Technologies: ICS/OT
View CWE Details
https://static.draeger.com/security
https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofed-network-m…