CVE-2026-0611
CRITICAL
9,2
Source: [email protected]
Attack Vector: network
Attack Complexity: low
Privileges Required: none
User Interaction: none
Confidentiality: N/A
Integrity: N/A
Availability: N/A
CRITICAL
9,8
Source: [email protected]
Attack Vector: network
Attack Complexity: low
Privileges Required: none
User Interaction: none
Scope: unchanged
Confidentiality: high
Integrity: high
Availability: high
Description
AI Translation Available
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
306
Missing Authentication for Critical Function
DraftCommon Consequences
Security Scopes Affected:
Access Control
Other
Potential Impacts:
Gain Privileges Or Assume Identity
Varies By Context
Applicable Platforms
Technologies:
Cloud Computing, ICS/OT
https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sen…
https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Sec…