CVE-2026-105667
HIGH
8,7
Source: [email protected]
Attack Vector: network
Attack Complexity: low
Privileges Required: none
User Interaction: none
Confidentiality: N/A
Integrity: N/A
Availability: N/A
HIGH
7,5
Source: [email protected]
Attack Vector: network
Attack Complexity: low
Privileges Required: none
User Interaction: none
Scope: unchanged
Confidentiality: none
Integrity: high
Availability: none
Description
AI Translation Available
In the default configuration, the HL7 listener accepts ADT^A40 patient-merge messages without authenticating the sending application. A remote sender can reassign studies and related workflow records from one patient identity to another. The archive's unmerge does not reverse this.
EPSS (Exploit Prediction Scoring System)
Trend Analysis
EPSS (Exploit Prediction Scoring System)
Prevede la probabilità di sfruttamento basata su intelligence sulle minacce e sulle caratteristiche della vulnerabilità.
EPSS Score
0,0021
Percentile
0,1th
Updated
EPSS Score Trend (Last 3 Days)
306
Missing Authentication for Critical Function
DraftCommon Consequences
Security Scopes Affected:
Access Control
Other
Potential Impacts:
Gain Privileges Or Assume Identity
Varies By Context
Applicable Platforms
Technologies:
Cloud Computing, ICS/OT
https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-…
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-288-02